UV-Visible/NIR Spectrophotometers Specifications

V-730 UV-Visible Spectrophotometer

Optical SystemRowland off-circle arrangement

Single monochromator

Double beam type
Light sourceHalogen lamp, Deuterium lamp
Wavelength range190 to 1,100 nm
Wavelength accuracy+/-0.2 nm (at 656.1 nm)
Wavelength repeatability+/-0.1 nm
Spectral bandwidth (SBW)1 nm
Stray light1 % (198 nm KCL 12 g/L aqueous solution)

0.02 % (220 nm NaI 10 g/L aqueous solution)

0.02 % (340 nm NaNO2 50 g/L aqueous solution)

0.02 % (370 nm NaNO2 50 g/L aqueous solution)

SBW: 1 nm
Photometric range-3~3 Abs
Photometric accuracy+/-0.0015 Abs (0 to 0.5 Abs)

+/-0.0025 Abs (0.5 to 1 Abs)

+/-0.3 %T Tested

with NIST SRM 930D
Photometric repeatability+/-0.0005 Abs (0 to 0.5 Abs)

+/-0.0005 Abs (0.5 to 1 Abs)

Tested with NIST SRM 930D
Scanning speed10-8000 nm/min
Slew speed24,000 nm/min
RMS noise0.00004 Abs

(0 Abs, wavelength: 500 nm, measurement time: 60 sec, SBW: 1 nm)
Baseline stability0.0004 Abs/hour

(Value obtained more than one hour after turning on the source,

when the room temperature is stabilized, wavelength: 250 nm,

response: slow)
Baseline flatness+/-0.0005 Abs (200 - 1,000 nm)
DetectorSilicon photodiode
Standard functionsIQ accessories, Start button, Analog output
Standard programsAbs/%T meter, Quantitative analysis, Spectrum measurement,

Time course measurement, Fixed wavelength measurement, Validation,

Daily maintenance,

Two wavelength time course measurement
Dimensions and weight486(W) x 441(D) x 216(H) mm, 15 kg
Power requirements120 VA
Installation requirementsRoom temperature: 15-30 Celsius, humidity: below 85%

V-750 UV-Visible Spectrophotometer

Optical SystemCzerny-Turner mount

Single monochromator

Fully symmetrical double beam type
Light SourceHalogen lamp, Deuterium lamp
Wavelength range190 to 900 nm
Wavelength accuracy+/-0.2 nm (at 656.1 nm)
Wavelength repeatability+/-0.05 nm
Spectral bandwidth (SBW)0.1, 0.2, 0.5, 1, 2, 5, 10 nm

L2, L5, L10 nm (low stray light mode)

M1, M2 nm (micro cell mode)
Stray light1 % (198 nm KCL 12 g/L aqueous solution)

0.005 % (220 nm NaI 10 g/L aqueous solution)

0.005 % (340 nm NaNO2 50 g/L aqueous solution)

0.005 % (370 nm NaNO2 50 g/L aqueous solution)

SBW: L2 nm
Photometric range-4~4 Abs
Photometric accuracy+/-0.0015 Abs (0 to 0.5 Abs)

+/-0.0025 Abs (0.5 to 1 Abs)

+/-0.3 %T

Tested with NIST SRM 930D
Photometric repeatability+/-0.0005 Abs (0 to 0.5 Abs)

+/-0.0005 Abs (0.5 to 1 Abs)

Tested with NIST SRM 930D
Scanning speed10~4000 nm/min (8,000 nm/min in preview mode)
Slew speed12,000 nm/min
RMS noise0.00003 Abs

(0 Abs, wavelength: 500 nm, measurement time: 60 sec, SBW: 2 nm)
Baseline stability0.0003 Abs/hour

(Wavelength: 250 nm, response: slow and SBW: 2 nm)
Baseline flatness+/-0.0002 Abs (200 - 850 nm)
DetectorPhotomultiplier tube
Standard functionsIQ Accessory, Start Button, Analog output
Standard programsAbs/%T meter, Quantitative analysis, Spectrum measurement,

Time course measurement, Fixed wavelength measurement, Validation,

Daily Check

Two wavelength time course measurement
Dimensions and weight460(W) x 602(D) x 268(H) mm, 27 kg
Power requirements150 VA
Installation requirementsRoom temperature: 15-30 Celsius, humidity: below 85%

V-760 UV-Visible Spectrophotometer

Optical SystemCzerny-Turner mount

Double monochromator

Fully symmetrical double beam type
Light SourceHalogen lamp, Deuterium lamp
Wavelength range187 to 900 nm
Wavelength accuracy+/-0.1 nm (at 656.1 nm)
Wavelength repeatability+/-0.05 nm
Spectral bandwidth (SBW)0.1, 0.2, 0.5, 1, 5, 2, 10 nm

L2, L5, L10 nm (low stray light mode)

M1, M2 nm (micro cell mode)
Stray light1 % (198 nm KCL 12 g/L aqueous solution)

0.00008 % (220 nm NaI 10 g/L aqueous solution)

0.00008 % (340 nm NaNO2 50 g/L aqueous solution)

0.00008 % (370 nm NaNO2 50 g/L aqueous solution)

SBW: L2 nm
Photometric range-4~6 Abs
Photometric accuracy+/-0.0015 Abs (0 to 0.5 Abs)

+/-0.0025 Abs (0.5 to 1 Abs)

+/-0.3 %T

Tested with NIST SRM 930D
Photometric repeatability+/-0.0005 Abs (0 to 0.5 Abs)

+/-0.0005 Abs (0.5 to 1 Abs)

Tested with NIST SRM 930D
Scanning speed10~4000 nm/min (8,000 nm/min in preview mode)
Slew speed12,000 nm/min
RMS noise0.00003 Abs

(0 Abs, wavelength: 500 nm, measurement time: 60 sec, SBW: 2 nm)
Baseline stability0.0003 Abs/hour

(Wavelength: 250 nm, response: slow

and SBW: 2 nm)
Baseline flatness+/-0.0003 Abs (200 - 800 nm)
DetectorPhotomultiplier tube
Standard functionsIQ Accessory, Start Button, Analog output
Standard programsAbs/%T meter, Quantitative analysis, Spectrum measurement,

Time course measurement, Fixed wavelength measurement, Validation,

Daily Check

Two wavelength time course measurement
Dimensions and Weight460(W) x 602(D) x 268(H) mm, 29 kg
Power requirements150 VA
Installation requirementsRoom temperature: 15-30 Celsius, humidity: below 85%

V-770 UV-Visible Spectrophotometer

Optical SystemCzerny-Turner grating mount
Single monochromator
Fully symmetrical double-beam
Light SourceHalogen lamp, Deuterium lamp
Wavelength range190 to 2,700 nm (3,200 nm, option)
Wavelength accuracy+/-0.3 nm (at 656.1 nm)
+/-1.5 nm (at 1,312.2 nm)
Wavelength repeatability+/-0.05 nm (UV-Vis), +/-0.2 nm (NIR)
Spectral bandwidth (SBW)UV-Visible: 0.1, 0.2, 0.5, 1, 2, 5, 10 nm
L2, L5, L10 nm (low stray light mode)
M1, M2 nm (micro cell mode)

NIR: 0.4, 0.8, 1, 2, 4, 8, 20, 40
L8, L20, L40 nm (low stray light mode)
M4, M8 nm (micro cell mode)
Stray light1 % (198 nm KCL)
0.005 % (220 nm NaI)
0.005 % (340 nm NaNO2)
0.005 % (370 nm NaNO2)
SBW: L2 nm

0.04 % (1420 nm: H2O)
0.1 % (1690 nm: CH2Br2)
SBW: L8 nm
Photometric rangeUV-Visible: -4~4 Abs
NIR: -3~3 Abs
Photometric accuracy+/-0.0015 Abs (0 to 0.5 Abs)
+/-0.0025 Abs (0.5 to 1 Abs)
+/-0.3 %T Tested with NIST SRM 930D
Photometric repeatability+/-0.0005 Abs (0 to 0.5 Abs)
+/-0.0005 Abs (0.5 to 1 Abs)
Tested with NIST SRM 930D
Scanning speed10~4,000 nm/min (8,000 nm/min in preview mode)
Slew speedUV-Vis: 12,000 nm/min
NIR: 48,000 nm/min
RMS noise0.00003 Abs
(0 Abs, wavelength: 500 nm, measurement time: 60 sec, SBW :2 nm)
Baseline stability0.0003 Abs/hour (Wavelength: 250 nm, response: slow and SBW: 2 nm)
Baseline flatness+/-0.0002 Abs (200 - 2,500 nm)
DetectorPMT, Peltier cooled PbS
Standard functionsIQ Accessory, Start Button, Analog output
Standard programsAbs/%T meter, Quantitative analysis, Spectrum measurement
Time course measurement, Fixed wavelength measurement
Validation, Daily check and Dual wavelength time course measurement
Dimensions and weight460(W) x 602(D) x 268(H) mm, 29 kg
Power requirements150 VA
Installation requirementsRoom temperature: 15-30 Celsius, humidity: below 85%

V-780 UV-Visible Spectrophotometer

Optical SystemCzerny-Turner mount

Single monochromator

Fully symmetrical double beam type
Light SourceHalogen lamp, Deuterium lamp
Wavelength range190 to 1,600 nm
Wavelength accuracy+/-0.3 nm (at 656.1 nm)

+/-1.0 nm (at 1312.2 nm)
Wavelength repeatability+/-0.05 nm (UV-Vis), +/-0.1 nm (NIR)
Spectral bandwidth (SBW)UV-Vis:

0.1, 0.2, 0.5, 1, 2, 5, 10 nm

L2, L5, L10 nm (low stray light mode)

M1, M2 nm (micro cell mode)

NIR:

0.2, 0.4, 0.5, 1, 2, 4, 10, 20,

L4, L10, L20 nm (low stray light mode)

M2, M4 nm (micro cell mode)
Stray light1 % (198 nm KCL 12 g/L aqueous solution)

0.005 % (220 nm NaI 10 g/L aqueous solution

0.005 % (340 nm NaNO2 50 g/L aqueous solution)

0.005 % (370 nm NaNO2 50 g/L aqueous solution)

SBW: L2nm


0.04 % (1420 nm: H2O)

SBW: L4 nm
Photometric rangeUV-Vis: -4~4 Abs

NIR: -3~3 Abs
Photometric accuracy+/-0.0015 Abs (0 to 0.5 Abs)

+/-0.0025 Abs (0.5 to 1 Abs)

+/-0.3 %T Tested

with NIST SRM 930D
Photometric repeatability+/-0.0005 Abs (0 to 0.5 Abs)

+/-0.0005 Abs (0.5 to 1 Abs)

Tested with NIST SRM 930D
Scanning speed10~4,000 nm/min (8,000 nm/min in preview mode)
Slew speedUV-Vis: 12000 nm/min

NIR: 24000 nm/min
RMS noise0.00003 Abs

(0 Abs, wavelength: 500 nm, measurement time: 60 sec, SBW:2 nm)
Baseline stability0.0003 Abs/hour

0.04 % (220 nm NaI 10 g/L aqueous solution)

0.02 % (340 nm NaNO2 50 g/L aqueous solution)

0.02 % (370 nm NaNO2 50 g/L aqueous solution)

(10 mm cell)
Baseline Stability±0.0004 Abs/hour

(Wavelength: 250 nm,

response: slow and SBW: 2nm.)
Baseline flatness+/-0.0002 Abs (200 - 1,600 nm)
DetectorPhotomultiplier tube, Peltier cooled InGaAs photodiode
Standard functionsIQ Accessory, Start Button, Analog output
Standard programsAbs/%T meter, Quantitative analysis, Spectrum measurement,

Time course measurement, Fixed wavelength measurement, Validation,

Daily Check

Two wavelength time course measurement
Dimensions and weight460(W) x 602(D) x 268(H) mm, 29 kg
Power requirements150 VA
Installation requirementsRoom temperature: 15-30 Celsius, humidity: below 85%