Probe Raman Specifications

MonochromatorSpectrograph; Czerny-Turner type monochromator f=200 mm
Wavelength ControlHigh accuracy rotary encoder Direct Drive
Wavelength Range (Raman Shift)100 - 8,000 cm-1
Resolution (cm-1/pixel)2.0 (standard)
Grating (Standard)900 gr/mm
Optional Gratings2,400, 1,800, 1,200, 600, 400, 300, 150 gr/mm (automated selection)
Number of GratingsMaximum 4 gratings on automated carousel
Detector1,650 x 200 (Andor BR246D CCD) or InGaAs Detector (1,064 nm)
Cooling Method4-stage Peltier (-60° C)
Laser wavelengthsRMP-510: 532 nm
RMP-520: 785 nm
RMP-530: 1,064 nm
Laser Power20 mW
AttenuatorManual: OPEN, OD4, OD1, OD0.3, CLOSED
Fiber ProbeLong Distance Objective: 20X (Option: 50X, 100X)
Beam Diameter20 µm (20X objective)
FiberSingle core, 1 m (Option: 3 m, 5 m, 10 m)
Sample ObservationCCD video camera
Observation Magnification250X (with 15 inch monitor, 20X objective, resolution 1,024 x 768)
Probe StageXYZ manual stage
X, Y ±4 mm, Z ±1.5 mm
Stage Size50 mm diameter
Laser SafetyClass 1; interlock system
Software/Operating SystemWindows 7, 8.1 or 10 Professional, Enterprise (32/64-bit )
Standard Software FunctionsSpectrum Measurement/Analysis, JASCO Canvas, Validation
Database Search FunctionOption - Bio-Rad Know-It-All® search (Spectrum search, structure search, User database, 600 compound standard database, AnalyzeIt™ (spectral interpretation)
Optional ProgramInterval Measurement Analysis
Dimension/Weight500 (W) x 360 (D) x 290 (H), 30 kg
Power RequirementsAC 100V +/- 10V, 200 VA