Analysis of Cloperastine Fendizoate listed in the Japanese Pharmacopoeia 18th Edition

April 21, 2026

Introduction

Cloperastine fendizoate, newly listed in the 18th edition of the Japanese Pharmacopoeia (JP18), is an antitussive agent used to relieve respiratory symptoms such as coughing by acting directly on the cough center. As with all pharmacopoeial drug substances, rigorous evaluation of purity is essential to ensure product quality, safety, and compliance with regulatory standards. The JP18 monograph defines specific criteria for system suitability and analytical performance in the purity testing of cloperastine fendizoate. Verification of these parameters is critical to confirm that the chromatographic system can provide accurate, precise, and reproducible results. In this application note, we present the results of system suitability testing for the purity analysis of cloperastine fendizoate using HPLC in accordance with JP18 requirements.

Experimental

Instruments

Pump:                                          PU-4580

Degassing unit:                          DG-4580

Low pressure gradient unit:     LG-4580

Autosampler:                              AS-4550

Column oven:                             CO-4060

UV detector:                               UV-4570

LC Conditions

Column:                                       Unifinepak C18 (4.6 mm I.D. x 150 mm L, 5 µm)

Eluent A:                                      0.1 M Potassium dihydrogen phosphate /acetonitrile/perchloric acid (400/320/1)

Eluent B:                                      Acetonitrile/0.1 M potassium dihydrogen phosphate/perchloric acid (1050/450/1)

Gradient:                                     A/B = 100/0 (0.0 min) → 100/0 (12.0 min) → 0/100 (22.0 min) → 0/100 (27.0 min) → 100/0 (27.1 min), 1 cycle; 37 min

Flow rate:                                    1.2 mL/min

Column temp:                            25 ºC

Wavelength:                               226 nm

Injection volume:                       20 µL

Standard:                                    4-Cholobenzophenone

(The dissolving and diluting solvent is eluent A)

Fig. 1 Structure of cloperastine fendizoate

Keywords

Japanese Pharmacopoeia, cloperastine fendizoate, UV detector, system suitability

Results

An overview of the system suitability requirements for the purity analysis of cloperastine fendizoate is summarized in Table 1.

Table 1. Overview of system suitability tests for the purity analysis of cloperastine fendizoate

TestCriteria
Detectability
When the standard solution (1.25 mg/L) is diluted to a concentration of 0.25 mg/L and 20 µL of the diluted standard solution (0.25 mg/L) is injected under the above operating conditions, the peak area of 4-chlorobenzophenone is 14% – 26% of the peak area of a 20 µL injection of the standard solution (1.25 mg/L).
System Performance
When 20 µL of the standard solution (1.25 mg/L) is injected under the above operating conditions, the number of theoretical plates and the peak symmetry factor for 4-chlorobenzophenone are ≥10,000 and ≤2.0, respectively.
System Reproducibility
When 20 µL of the standard solution (1.25 mg/L) is injected 6 times under the above operating conditions, the relative standard deviation of the 4-chlorobenzophenone peak area is ≤2.0 %.

Figure 2 shows the results of the purity detectability test. The ratio of the 4-chlorobenzophenone peak area for the diluted solution (0.25 mg/L) relative to the standard solution (1.25 mg/L) was 19.5%, which met the specified criterion of 14% – 26 %.

Fig. 2 Chromatograms of 4-chlorobenzophenone for the test of the detectability
(left: 1.25 mg/L, right: 0.25 mg/L)

Figure 3 shows chromatograms of 4-chlorobenzophenone standard solutions (n = 6), and Table 2 summarizes the peak area reproducibility results based on system performance and reproducibility. The measured number of theoretical plates was 80,288 (criterion: ≥10,000), the symmetry factor was 1.01 (criterion: ≤2.0), and the relative standard deviation of the peak area was 0.21 % (criterion: ≤2.0 %). All parameters met the specified system suitability criteria.

Fig. 3 Chromatogram of 4-chlorobenzophenone standard solution
(1.25 mg/L, n = 6)

Table 2. 4-Chlorobenzophenone peak area reproducibility (n = 6)

Injection Number
Peak Area
133,336
233,528
333,513
433,485
533,450
633,443
Average
33,459
SD69
RSD [%]
0.21

Conclusion

System suitability tests for the purity analysis of cloperastine fendizoate, which is newly listed in the 18th edition of the Japanese Pharmacopoeia (JP18), were successfully performed. As shown in Table 3, all evaluated parameters met the specified acceptance criteria of the Japanese Pharmacopoeia.

Table 3. System suitability criterion and results for the purity of cloperastine fendizoate

TestTest ItemCriterionResultJudgement
Detectability
Area ratio
14–26 %
19.5 %
Passed
System Performance
Number of theoretical plates
≥10,000
80,288
Passed
System Performance
Symmetry factor
≤2.0
1.01
Passed
System Reproducibility
Relative standard deviation of peak area
≤2.0 %
0.21 %
Passed
This document has been prepared based on information available at the time of publication and is subject to revision without notice. Although the contents are checked with the utmost care, we do not guarantee their accuracy or completeness. JASCO Corporation assumes no responsibility or liability for any loss or damage incurred as a result of the use of any information contained in this document. Copyright and other intellectual property rights in this document remain the property of JASCO Corporation. Please do not attempt to copy, modify, redistribute, or sell etc. in whole or in part without prior written permission.