Skip to content

JASCO JASCO

  • News
  • Events
  • E-Store
  • My Account
  • Contact Us
  • Worldwide
Search
Click to view menu
MENUMENU
  • Products
    • Chromatography
      • HPLC
      • RHPLC
      • UHPLC
      • LC-MS
      • Preparative LC
      • Analytical SFC
      • Semi-Preparative SFC
      • Hybrid SFC
      • Fuel Analysis by SFC-FID
      • Preparative SFC
      • Supercritical Fluid Extraction
      • Chromatography Software
    • Molecular Spectroscopy
      • Circular Dichroism
      • High-Throughput CD
      • Vibrational CD
      • Circularly Polarized Luminescence (CPL)
      • Polarimeters
      • FTIR Spectrometers
      • FTIR Microscopy
      • FTIR Portable
      • Raman Microscopy
      • Palmtop Raman Spectrometer
      • Probe Raman
      • UV-Visible/NIR Spectrophotometers
      • UV-Visible/NIR Microscopy
      • Fluorescence Spectrophotometers
      • Film Thickness
      • Spectra Manager™ Suite
    • Refurbished
      • Refurbished HPLC Systems
      • HPLC Switching Valves
      • FTIR Accessories
  • Service
    • Service and Support Plans
    • Service Request Form
  • Applications
  • KnowledgeBase
  • Learning Center
    • Best Practice
      • Circular Dichroism Tips & Tricks for Biological Samples
      • CD Scale Calibration with ACS
      • Fluorescence Tips & Tricks
      • Raman Spectroscopy Tips & Tricks
    • Training Videos
      • ChromNAV
      • SF-NAV
      • Circular Dichroism
      • UV-Visible/NIR
      • Fluorescence
    • Training Seminars
      • Training Registration Form
    • Webinars
    • eBooks
    • Theory
      • Theory of Molecular Spectroscopy
      • Chromatography
  • About Us
    • President’s Message
    • Contact
    • History
    • Careers
  • News
  • Events
  • Worldwide
  • Shop
  • My Account
  • Contact Us

Home / Applications / Avoiding Interference from Fluorescence Signal by Performing Raman Spectroscopy Measurements at Excitation Wavelength of 1064 nm

  • Industry

  • Technique

Avoiding Interference from Fluorescence Signal by Performing Raman Spectroscopy Measurements at Excitation Wavelength of 1064 nm

PDF IconDownload This Application

June 12, 2025

Introduction

In Raman spectroscopy, an excitation wavelength of 532 nm is often chosen due to the higher sensitivity of the detector at this wavelength and the higher intensity of the Raman scattered light. However, at this excitation wavelength, fluorescence emitted from the sample can increase the baseline, posing a challenge for analysis¹). One known method to address this issue is to increase the excitation wavelength to avoid fluorescence from the sample2). Since the intensity of Raman scattered light is approximately inversely proportional to the fourth power of the excitation wavelength³), the use of a longer excitation wavelength results in a decrease in the Raman scattering intensity.

In the experiment reported in this study, a Raman microscope equipped with a 1064 nm excitation wavelength laser was used to measure two types of fluorescent samples that are difficult to analyze with excitation wavelengths in the visible region, such as 532 nm. The effectiveness of this approach was further confirmed by performing Raman measurements on microscopic plastic particles. The instrument used in this study is equipped with various optical components, including a detector and objective lenses optimized for the near-infrared region. As a result, the reduction in Raman scattering intensity due to the use of a long-wavelength laser is mitigated.

Experimental

Fig. 1 PEEK plastic
Fig. 2 Cassis liqueur

Samples

  1. Polyether ether ketone (PEEK) plastic (Fig.1)
  2. Cassis liqueur (Fig.2)
  3. Polystyrene (PS) particles (⌀2 µm)

System

Instrument: NRS-4500 Raman Spectrometer (Fig.3)

Parameters

  1. Ex wavelength:
    • 1064 nm
  2. Objective lens:
    • 20x for near-infrared region (for PEEK and cassis liqueur)
    • 100x for near-infrared region (for ⌀2 µm PS particles)
  3. Rejection filter:
    • 1064 nm long-pass edge filter
  4. Detector:
    • InGaAs detector

      NRS-4500 Raman Microscope
      Fig. 3 NRS-4500 Raman spectrometer

*Other than the parameters above, the measurements were conducted using optimal parameters for each sample.

 

Keywords

Raman spectroscopy, reduction of fluorescence interference, near-infrared excitation, 1064 nm laser, Raman Microscope

Results

  • Fluorescent samples:
Fig. 4 Raman spectra of PEEK for different excitation wavelengths

Figure 4 shows the Raman spectrum of PEEK, which is a engineering plastic, measured at an excitation wavelength of 1064 nm. For comparison, spectra at excitation wavelengths of 532 and 785 nm are also shown. In the spectrum excited at 532 nm, the Raman peaks associated with the molecular structure are almost undetectable due to the strong fluorescence emitted from the sample, while in the spectrum excited at 785 nm, the fluorescence is reduced and Raman peaks are weakly observed. However, if an excitation wavelength of 1064 nm is used, baseline elevation due to fluorescence is strongly suppressed, resulting in much clearer Raman peaks.

As shown in Figure 5, the use of an excitation wavelength of 1064 nm allows the peaks due to different functional groups in PEEK to be easily assigned. Thus, changing the excitation wavelength from the visible to the near-infrared region reduces sample fluorescence, allowing a qualitative analysis of the Raman spectrum.

Fig. 5 Raman peak assignment for PEEK

Fig. 6 Raman spectra of cassis liqueur for different excitation wavelengths

Figure 6 shows the Raman spectrum of a colored liquid (cassis liqueur) measured at an excitation wavelength of 1064 nm. As in Figure 4, spectra at excitation wavelengths of 532 nm and 785 nm are also shown for comparison. No clear peaks are seen in the spectrum excited at 532 nm, nor in that excited at 785 nm, a wavelength commonly used to avoid sample fluorescence. However, the spectrum excited at 1064 nm exhibits clear peaks.

As shown in Figure 7, each peak can be easily assigned, and those corresponding to ethanol and sugar can be identified. Thus, even for samples where useful Raman spectra cannot be obtained by excitation at 785 nm, high-quality spectra can be acquired using an excitation wavelength of 1064 nm.

Fig. 7 Raman peak assignment for cassis liqueur

  • Plastic Particles:
Fig. 8 PS image and Raman spectrum

Figure 8 shows the measurement results for polystyrene particles with a diameter of 2 µm using an excitation wavelength of 1064 nm, and Figure 9 presents the results of a database search for the obtained Raman spectrum. From Figures 8 and 9, it can be seen that the obtained Raman spectrum is of sufficient quality for qualitative analysis. The theoretical laser spot size at a wavelength of 1064 nm is approximately 2 µm. Even when measuring plastic particles with a size comparable to this laser spot size, qualitative analysis using Raman spectroscopy is possible.

Fig. 9 Database search results for Raman spectrum

Conclusion

Using a JASCO Raman microscope equipped with a laser that allows excitation at 1064 nm suppresses sample fluorescence, making it possible to measure samples that are difficult to analyze with excitation wavelengths in the visible region. Additionally, qualitative analysis of microscopic particles using Raman spectroscopy was shown to be applicable to microscopic measurements. Since excitation at 1064 nm allows greater suppression of fluorescence than visible excitation, this analytical method can be useful for analyzing highly fluorescent samples such as pharmaceuticals, pigmented materials, and biological materials.

References

1. H. Kawasaki, Y. Nakahara, T. Hasegawa (Eds): “Handbook of Instrumental Analysis. 1”, p. 163 (2020), (Kagaku-Dojin Publishing Company, INC)

2. M. Tasumi (Ed): “Infrared Spectroscopy –Basics and Latest Methods”, The Spectroscopical Society of Japan Edition, p. 185 (2012), (ST Japan Inc.)

3. H. Hamaguchi, K Iwata (Eds): “Raman Spectroscopy”, P.53 (2015), KODANSHA SCIENTIFIC LTD.

This document has been prepared based on information available at the time of publication and is subject to revision without notice. Although the contents are checked with the utmost care, we do not guarantee their accuracy or completeness. JASCO Corporation assumes no responsibility or liability for any loss or damage incurred as a result of the use of any information contained in this document. Copyright and other intellectual property rights in this document remain the property of JASCO Corporation. Please do not attempt to copy, modify, redistribute, or sell etc. in whole or in part without prior written permission.
JASCO Application Note

Avoiding Interference from Fluorescence Signal by Performing Raman Spectroscopy Measurements at Excitation Wavelength of 1064 nm

Introduction

In Raman spectroscopy, an excitation wavelength of 532 nm is often chosen due to the higher sensitivity of the detector at this wavelength and the higher intensity of the Raman scattered light. However, at this excitation wavelength, fluorescence emitted from the sample can increase the baseline, posing a challenge for analysis¹). One known method to address this issue is to increase the excitation wavelength to avoid fluorescence from the sample2). Since the intensity of Raman scattered light is approximately inversely proportional to the fourth power of the excitation wavelength³), the use of a longer excitation wavelength results in a decrease in the Raman scattering intensity.

In the experiment reported in this study, a Raman microscope equipped with a 1064 nm excitation wavelength laser was used to measure two types of fluorescent samples that are difficult to analyze with excitation wavelengths in the visible region, such as 532 nm. The effectiveness of this approach was further confirmed by performing Raman measurements on microscopic plastic particles. The instrument used in this study is equipped with various optical components, including a detector and objective lenses optimized for the near-infrared region. As a result, the reduction in Raman scattering intensity due to the use of a long-wavelength laser is mitigated.

Experimental

Fig. 1 PEEK plastic
Fig. 2 Cassis liqueur

Samples

  1. Polyether ether ketone (PEEK) plastic (Fig.1)
  2. Cassis liqueur (Fig.2)
  3. Polystyrene (PS) particles (⌀2 µm)

System

Instrument: NRS-4500 Raman Spectrometer (Fig.3)

Parameters

  1. Ex wavelength:
    • 1064 nm
  2. Objective lens:
    • 20x for near-infrared region (for PEEK and cassis liqueur)
    • 100x for near-infrared region (for ⌀2 µm PS particles)
  3. Rejection filter:
    • 1064 nm long-pass edge filter
  4. Detector:
    • InGaAs detector

      NRS-4500 Raman Microscope
      Fig. 3 NRS-4500 Raman spectrometer

*Other than the parameters above, the measurements were conducted using optimal parameters for each sample.

 

Results

  • Fluorescent samples:
Fig. 4 Raman spectra of PEEK for different excitation wavelengths

Figure 4 shows the Raman spectrum of PEEK, which is a engineering plastic, measured at an excitation wavelength of 1064 nm. For comparison, spectra at excitation wavelengths of 532 and 785 nm are also shown. In the spectrum excited at 532 nm, the Raman peaks associated with the molecular structure are almost undetectable due to the strong fluorescence emitted from the sample, while in the spectrum excited at 785 nm, the fluorescence is reduced and Raman peaks are weakly observed. However, if an excitation wavelength of 1064 nm is used, baseline elevation due to fluorescence is strongly suppressed, resulting in much clearer Raman peaks.

As shown in Figure 5, the use of an excitation wavelength of 1064 nm allows the peaks due to different functional groups in PEEK to be easily assigned. Thus, changing the excitation wavelength from the visible to the near-infrared region reduces sample fluorescence, allowing a qualitative analysis of the Raman spectrum.

Fig. 5 Raman peak assignment for PEEK

Fig. 6 Raman spectra of cassis liqueur for different excitation wavelengths

Figure 6 shows the Raman spectrum of a colored liquid (cassis liqueur) measured at an excitation wavelength of 1064 nm. As in Figure 4, spectra at excitation wavelengths of 532 nm and 785 nm are also shown for comparison. No clear peaks are seen in the spectrum excited at 532 nm, nor in that excited at 785 nm, a wavelength commonly used to avoid sample fluorescence. However, the spectrum excited at 1064 nm exhibits clear peaks.

As shown in Figure 7, each peak can be easily assigned, and those corresponding to ethanol and sugar can be identified. Thus, even for samples where useful Raman spectra cannot be obtained by excitation at 785 nm, high-quality spectra can be acquired using an excitation wavelength of 1064 nm.

Fig. 7 Raman peak assignment for cassis liqueur

  • Plastic Particles:
Fig. 8 PS image and Raman spectrum

Figure 8 shows the measurement results for polystyrene particles with a diameter of 2 µm using an excitation wavelength of 1064 nm, and Figure 9 presents the results of a database search for the obtained Raman spectrum. From Figures 8 and 9, it can be seen that the obtained Raman spectrum is of sufficient quality for qualitative analysis. The theoretical laser spot size at a wavelength of 1064 nm is approximately 2 µm. Even when measuring plastic particles with a size comparable to this laser spot size, qualitative analysis using Raman spectroscopy is possible.

Fig. 9 Database search results for Raman spectrum

Conclusion

Using a JASCO Raman microscope equipped with a laser that allows excitation at 1064 nm suppresses sample fluorescence, making it possible to measure samples that are difficult to analyze with excitation wavelengths in the visible region. Additionally, qualitative analysis of microscopic particles using Raman spectroscopy was shown to be applicable to microscopic measurements. Since excitation at 1064 nm allows greater suppression of fluorescence than visible excitation, this analytical method can be useful for analyzing highly fluorescent samples such as pharmaceuticals, pigmented materials, and biological materials.

Keywords

Raman spectroscopy, reduction of fluorescence interference, near-infrared excitation, 1064 nm laser, Raman Microscope

References

1. H. Kawasaki, Y. Nakahara, T. Hasegawa (Eds): “Handbook of Instrumental Analysis. 1”, p. 163 (2020), (Kagaku-Dojin Publishing Company, INC)

2. M. Tasumi (Ed): “Infrared Spectroscopy –Basics and Latest Methods”, The Spectroscopical Society of Japan Edition, p. 185 (2012), (ST Japan Inc.)

3. H. Hamaguchi, K Iwata (Eds): “Raman Spectroscopy”, P.53 (2015), KODANSHA SCIENTIFIC LTD.

This document has been prepared based on information available at the time of publication and is subject to revision without notice. Although the contents are checked with the utmost care, we do not guarantee their accuracy or completeness. JASCO Corporation assumes no responsibility or liability for any loss or damage incurred as a result of the use of any information contained in this document. Copyright and other intellectual property rights in this document remain the property of JASCO Corporation. Please do not attempt to copy, modify, redistribute, or sell etc. in whole or in part without prior written permission.
28600 Mary’s Court, Easton, MD 21601 USA • (800) 333-5272 • Fax: (410) 822-7526 • jascoinc.com/applications

Close

Designed in Tokyo. TRUSTED globally.

View our support plans

Connect with JASCO

  • Facebook
  • Twitter
  • LinkedIn
  • JASCO Sales
  • 800-333-5272

Receive the latest promotions and special offers

  • This field is for validation purposes and should be left unchanged.
  • Careers
  • Press Kit
  • JASCO Privacy Policy
  • Sitemap
  • Environmental Policy

© , JASCO. All Rights Reserved.