Skip to content

JASCO JASCO

  • News
  • Events
  • E-Store
  • My Account
  • Contact Us
  • Worldwide
Search
Click to view menu
MENUMENU
  • Products
    • Chromatography
      • HPLC
      • RHPLC
      • UHPLC
      • LC-MS
      • Preparative LC
      • Analytical SFC
      • Semi-Preparative SFC
      • Hybrid SFC
      • Fuel Analysis by SFC-FID
      • Preparative SFC
      • Supercritical Fluid Extraction
      • Chromatography Software
    • Molecular Spectroscopy
      • Circular Dichroism
      • High-Throughput CD
      • Vibrational CD
      • Circularly Polarized Luminescence (CPL)
      • Polarimeters
      • FTIR Spectrometers
      • FTIR Microscopy
      • FTIR Portable
      • Raman Microscopy
      • Palmtop Raman Spectrometer
      • Probe Raman
      • UV-Visible/NIR Spectrophotometers
      • UV-Visible/NIR Microscopy
      • Fluorescence Spectrophotometers
      • Film Thickness
      • Spectra Manager™ Suite
    • Refurbished
      • Refurbished HPLC Systems
      • HPLC Switching Valves
      • FTIR Accessories
  • Service
    • Service and Support Plans
    • Service Request Form
  • Applications
  • KnowledgeBase
  • Learning Center
    • Best Practice
      • Circular Dichroism Tips & Tricks for Biological Samples
      • CD Scale Calibration with ACS
      • Fluorescence Tips & Tricks
      • Raman Spectroscopy Tips & Tricks
    • Training Videos
      • ChromNAV
      • SF-NAV
      • Circular Dichroism
      • UV-Visible/NIR
      • Fluorescence
    • Training Seminars
      • Training Registration Form
    • Webinars
    • eBooks
    • Theory
      • Theory of Molecular Spectroscopy
      • Chromatography
  • About Us
    • President’s Message
    • Contact
    • History
    • Careers
  • News
  • Events
  • Worldwide
  • Shop
  • My Account
  • Contact Us

Home / Applications / Use of Diamond Compression Cell for Different Sample Types

  • Industry

  • Technique

Use of Diamond Compression Cell for Different Sample Types

PDF IconDownload This Application

April 21, 2025

Introduction

Transmission measurements are often used in infrared (IR) spectroscopy for qualitative analysis of foreign substances because of the extensive spectral database that is available. However, the surface condition and thickness of the sample must be properly controlled in order to achieve precise measurements. Pretreatment often involves thinning samples using a microtome or slicer, or compressing samples using a press, but these methods require a certain level of proficiency. Therefore, there is a need to simplify this complicated pretreatment process.

Fig. 1 DC-500 Diamond compression cell

For this purpose, JASCO offers the DC-500 diamond compression cell (Figure 1). In this cell, two diamond windows are placed in the centers of a base plate and a pressure plate (Figure 2). The sample is set on the base-plate window, and is then covered by the pressure-plate window to form a sandwich structure. Pressure is then applied using two thumbscrews to compress the sample (Figure 3). This cell makes it possible to measure the transmittance of a sample without the need for pretreatment using a slicer or pressing equipment. Sample preparation does not require any specialized skill, since sample immobilization and compression are performed using only two thumbscrews. This cell can be easily cleaned with organic solvents such as ethanol or acetone because no adhesive is used between the stainless steel and diamond window plates.

Fig. 2 DC-500 Diamond compression cell
(Enlarged view of area around diamond window of base plate)

Generally, before performing measurements, the pressure plate with the diamond window is removed after compressing the sample. This is to avoid the occurrence of interference fringes in the IR spectrum due to the gap between the two diamond plates. However, for some types of sample (e.g., rubber-like samples), it is better not to remove the pressure plate, as described later. In this study, three types of sample were prepared for microscopic FT-IR measurements, and the effect of the number of diamond windows used on the qualitative analysis results was investigated.

Fig. 3 Procedure for using diamond compression cell

Specifications of DC-500 Diamond Compression Cell
Diamond window plate:

⌀2.5 mm (diameter for which analysis is valid: ⌀2.0 mm)
Thickness 1.0 mm
Pressurization:
Hand-cranking using thumbscrews
Diamond properties:
Refractive index 2.4
Insoluble in water
Acid and base-resistantNote 1
High mechanical strengthNote 2
Measurement target:  General samples
Acid, solvent, and oxidation catalyst samples can also be usedNote 1

Note 1:  Although diamond itself can be used for strongly acidic or strongly basic samples, the surrounding metal (stainless steel) holding the diamond plate may corrode. After the measurement, remove the sample and clean the diamond cell immediately.

Note 2:  If a strong force is applied to the diamond windows causing them to make contact with each other, the windows may chip.

Experimental

Sample
Ethylene-vinyl acetate: Transparent resin
Polyisoprene:  Natural rubber
Calcium carbonate:  Inorganic powder

System
Instrument:   FT/IR-4X FTIR spectrometer
Accessories:  IRT-5200 IR microscope
DC-500 Diamond compression cell

Parameters
Resolution:  4 cm-1
Objective lens:  Cassegrain objective 16x
Detector:  Mid-band MCT detector
Measurement size:  30 µm x 30 µm

Keywords

Diamond compression cell, DC-500, foreign substance analysis, transmission measurement, pretreatment, FT-IR, FT-IR microscopy

Results

Ethylene-Vinyl Acetate

Fig. 4 IR Spectrum

A transparent resin, ethylene-vinyl acetate, was measured using either one or two diamond plates (Figure 4). In the case of two diamond plates, interference fringes are observed in the baseline region (blue curve). These fringes are due to light interfering in the gap between the plates. However, it was found that for this sample, removing the top diamond plate did not cause the sample shape to change, and so measurements could be performed using only the bottom plate. The resulting spectrum is seen to be free from interference fringes (fuchsia curve). Thus, a single diamond plate is suitable for measurement of samples whose shape is maintained after compression.

Fig. 5 KnowItAll search results

A search was performed using KnowItAll (John Wiley & Sons, Inc.) for the IR spectrum obtained using two diamond plates, in which interference fringes appear (Figure 5). Since the sample was correctly identified, qualitative analysis is possible even when interference fringes are present if the height of the main peaks is sufficiently large. However, for quantitative analysis where accurately determining peak heights is important, care must be taken because interference fringes can be a major problem.

Polyisoprene

Fig. 6 IR spectrum

A natural rubber, polyisoprene, was next measured. The results in Figure 6 show that a high-quality spectrum was obtained using two diamond plates, whereas peak saturation occurred when a single diamond plate was used. Interference fringes, which are clearly seen in Figure 4, are barely visible in Figure 6. Polyisoprene is an elastic sample that will return to its original shape unless it is continuously compressed. For such samples, measurement using two diamond plates is therefore more suitable.

Calcium Carbonate

Fig. 7 FTIR spectrum

Figure 7 shows IR spectra of an inorganic powder, calcium carbonate, obtained using one or two diamond plates. Although interference fringes appear in the case of two diamond plates, their intensity is much lower than in Figure 4. This is thought to be the result of light scattering by the powder sample and the non-parallel state of the diamond plates due to the hardness of the sample, which reduce the degree of interference. Thus, for powder samples that retain their shape after compression, strong interference fringes are unlikely to appear in the spectrum, and so measurements with either one or two diamond plates are possible.

Fig. 8 KnowItAll search results

Figure 8 shows the results of a KnowItAll search for the IR spectrum containing interference fringes. Again, the sample was correctly identified as calcium carbonate, indicating that qualitative analysis is possible using two diamond plates. For samples such as calcium carbonate that do not adhere to the diamond plate after compression, the use of two diamond plates may be preferable for easier sample handling.

Conclusion

Table 1 shows the recommended use of the diamond compression cell for three different types of sample based on the results of this study. Cases are shown for which it is more appropriate to use one diamond plate due to the influence of interference fringes, when it is better to use two diamond plates due to sample elasticity, and when either method can be used due to the shape and hardness of the sample.

The JASCO diamond compression cell provides a simple pretreatment method for acquisition of accurate IR spectra of samples with a variety of properties, such as resins, inorganic powder, fibers, and rubber.

Table 1   Use of diamond compression cell for different samples

SampleOne PlateTwo PlatesExample
The sample shape is retained even if pressure is not continually applied.OptimalNot recommended in some casesResins, saccharides, single fibers, etc.
The sample shape is not retained unless pressure is continually applied.UnsuitableOptimalPolyisoprene (rubber, etc.)
The sample shape is retained even if pressure is not applied continuously, but the stretched surfaces are not parallel or light is scattered by the sample.OptimalSuitableInorganic powder, fibers

 

We thank Furukawa Electric Co., Ltd. for providing some of the samples and for guidance on the measurement methods used in this report.

 

Applicative Solution Lab Division  Y. Higuchi

This document has been prepared based on information available at the time of publication and is subject to revision without notice. Although the contents are checked with the utmost care, we do not guarantee their accuracy or completeness. JASCO Corporation assumes no responsibility or liability for any loss or damage incurred as a result of the use of any information contained in this document. Copyright and other intellectual property rights in this document remain the property of JASCO Corporation. Please do not attempt to copy, modify, redistribute, or sell etc. in whole or in part without prior written permission.
JASCO Application Note

Use of Diamond Compression Cell for Different Sample Types

Introduction

Transmission measurements are often used in infrared (IR) spectroscopy for qualitative analysis of foreign substances because of the extensive spectral database that is available. However, the surface condition and thickness of the sample must be properly controlled in order to achieve precise measurements. Pretreatment often involves thinning samples using a microtome or slicer, or compressing samples using a press, but these methods require a certain level of proficiency. Therefore, there is a need to simplify this complicated pretreatment process.

Fig. 1 DC-500 Diamond compression cell

For this purpose, JASCO offers the DC-500 diamond compression cell (Figure 1). In this cell, two diamond windows are placed in the centers of a base plate and a pressure plate (Figure 2). The sample is set on the base-plate window, and is then covered by the pressure-plate window to form a sandwich structure. Pressure is then applied using two thumbscrews to compress the sample (Figure 3). This cell makes it possible to measure the transmittance of a sample without the need for pretreatment using a slicer or pressing equipment. Sample preparation does not require any specialized skill, since sample immobilization and compression are performed using only two thumbscrews. This cell can be easily cleaned with organic solvents such as ethanol or acetone because no adhesive is used between the stainless steel and diamond window plates.

Fig. 2 DC-500 Diamond compression cell
(Enlarged view of area around diamond window of base plate)

Generally, before performing measurements, the pressure plate with the diamond window is removed after compressing the sample. This is to avoid the occurrence of interference fringes in the IR spectrum due to the gap between the two diamond plates. However, for some types of sample (e.g., rubber-like samples), it is better not to remove the pressure plate, as described later. In this study, three types of sample were prepared for microscopic FT-IR measurements, and the effect of the number of diamond windows used on the qualitative analysis results was investigated.

Fig. 3 Procedure for using diamond compression cell

Specifications of DC-500 Diamond Compression Cell
Diamond window plate:

⌀2.5 mm (diameter for which analysis is valid: ⌀2.0 mm)
Thickness 1.0 mm
Pressurization:
Hand-cranking using thumbscrews
Diamond properties:
Refractive index 2.4
Insoluble in water
Acid and base-resistantNote 1
High mechanical strengthNote 2
Measurement target:  General samples
Acid, solvent, and oxidation catalyst samples can also be usedNote 1

Note 1:  Although diamond itself can be used for strongly acidic or strongly basic samples, the surrounding metal (stainless steel) holding the diamond plate may corrode. After the measurement, remove the sample and clean the diamond cell immediately.

Note 2:  If a strong force is applied to the diamond windows causing them to make contact with each other, the windows may chip.

Experimental

Sample
Ethylene-vinyl acetate: Transparent resin
Polyisoprene:  Natural rubber
Calcium carbonate:  Inorganic powder

System
Instrument:   FT/IR-4X FTIR spectrometer
Accessories:  IRT-5200 IR microscope
DC-500 Diamond compression cell

Parameters
Resolution:  4 cm-1
Objective lens:  Cassegrain objective 16x
Detector:  Mid-band MCT detector
Measurement size:  30 µm x 30 µm

Results

Ethylene-Vinyl Acetate

Fig. 4 IR Spectrum

A transparent resin, ethylene-vinyl acetate, was measured using either one or two diamond plates (Figure 4). In the case of two diamond plates, interference fringes are observed in the baseline region (blue curve). These fringes are due to light interfering in the gap between the plates. However, it was found that for this sample, removing the top diamond plate did not cause the sample shape to change, and so measurements could be performed using only the bottom plate. The resulting spectrum is seen to be free from interference fringes (fuchsia curve). Thus, a single diamond plate is suitable for measurement of samples whose shape is maintained after compression.

Fig. 5 KnowItAll search results

A search was performed using KnowItAll (John Wiley & Sons, Inc.) for the IR spectrum obtained using two diamond plates, in which interference fringes appear (Figure 5). Since the sample was correctly identified, qualitative analysis is possible even when interference fringes are present if the height of the main peaks is sufficiently large. However, for quantitative analysis where accurately determining peak heights is important, care must be taken because interference fringes can be a major problem.

Polyisoprene

Fig. 6 IR spectrum

A natural rubber, polyisoprene, was next measured. The results in Figure 6 show that a high-quality spectrum was obtained using two diamond plates, whereas peak saturation occurred when a single diamond plate was used. Interference fringes, which are clearly seen in Figure 4, are barely visible in Figure 6. Polyisoprene is an elastic sample that will return to its original shape unless it is continuously compressed. For such samples, measurement using two diamond plates is therefore more suitable.

Calcium Carbonate

Fig. 7 FTIR spectrum

Figure 7 shows IR spectra of an inorganic powder, calcium carbonate, obtained using one or two diamond plates. Although interference fringes appear in the case of two diamond plates, their intensity is much lower than in Figure 4. This is thought to be the result of light scattering by the powder sample and the non-parallel state of the diamond plates due to the hardness of the sample, which reduce the degree of interference. Thus, for powder samples that retain their shape after compression, strong interference fringes are unlikely to appear in the spectrum, and so measurements with either one or two diamond plates are possible.

Fig. 8 KnowItAll search results

Figure 8 shows the results of a KnowItAll search for the IR spectrum containing interference fringes. Again, the sample was correctly identified as calcium carbonate, indicating that qualitative analysis is possible using two diamond plates. For samples such as calcium carbonate that do not adhere to the diamond plate after compression, the use of two diamond plates may be preferable for easier sample handling.

Conclusion

Table 1 shows the recommended use of the diamond compression cell for three different types of sample based on the results of this study. Cases are shown for which it is more appropriate to use one diamond plate due to the influence of interference fringes, when it is better to use two diamond plates due to sample elasticity, and when either method can be used due to the shape and hardness of the sample.

The JASCO diamond compression cell provides a simple pretreatment method for acquisition of accurate IR spectra of samples with a variety of properties, such as resins, inorganic powder, fibers, and rubber.

Table 1   Use of diamond compression cell for different samples

SampleOne PlateTwo PlatesExample
The sample shape is retained even if pressure is not continually applied.OptimalNot recommended in some casesResins, saccharides, single fibers, etc.
The sample shape is not retained unless pressure is continually applied.UnsuitableOptimalPolyisoprene (rubber, etc.)
The sample shape is retained even if pressure is not applied continuously, but the stretched surfaces are not parallel or light is scattered by the sample.OptimalSuitableInorganic powder, fibers

 

We thank Furukawa Electric Co., Ltd. for providing some of the samples and for guidance on the measurement methods used in this report.

 

Applicative Solution Lab Division  Y. Higuchi

Keywords

Diamond compression cell, DC-500, foreign substance analysis, transmission measurement, pretreatment, FT-IR, FT-IR microscopy

This document has been prepared based on information available at the time of publication and is subject to revision without notice. Although the contents are checked with the utmost care, we do not guarantee their accuracy or completeness. JASCO Corporation assumes no responsibility or liability for any loss or damage incurred as a result of the use of any information contained in this document. Copyright and other intellectual property rights in this document remain the property of JASCO Corporation. Please do not attempt to copy, modify, redistribute, or sell etc. in whole or in part without prior written permission.
28600 Mary’s Court, Easton, MD 21601 USA • (800) 333-5272 • Fax: (410) 822-7526 • jascoinc.com/applications

Close

Designed in Tokyo. TRUSTED globally.

View our support plans

Connect with JASCO

  • Facebook
  • Twitter
  • LinkedIn
  • JASCO Sales
  • 800-333-5272

Receive the latest promotions and special offers

  • This field is for validation purposes and should be left unchanged.
  • Careers
  • Press Kit
  • JASCO Privacy Policy
  • Sitemap
  • Environmental Policy

© , JASCO. All Rights Reserved.