Choosing the Right Cuvette Material
Material determines what you can measure. Special optical glass transmits from 320 to 2,500 nanometers and costs less, making it the right call for visible-only work. Spectrosil Quartz transmits from 190 to 2,700 nanometers, covering UV, visible, and near-infrared in a single cell. If any part of your method dips below 320 nanometers, quartz is not optional.
| Material | Wavelength Range | Best For |
|---|---|---|
| Special optical glass | 320–2,500 nm | Visible-only measurements where cost matters |
| Spectrosil Quartz | 190–2,700 nm | UV, visible, and NIR measurements in one cell |
For a full walkthrough of material selection by application, see Tips and Tricks for UV-Visible Spectroscopy.
Rectangular Glass Cells

Special optical glass, 320 to 2,500 nanometers. The standard choice for routine visible-range work.
| Part Number | Description |
|---|---|
| 1103-4011 | Rectangular glass cell 10 x 1 Special optical glass with lid 320 -2500 nm |
| 1103-4012 | Rectangular glass cell 10 x 2 Special optical glass with lid 320 -2500 nm |
| 1103-4013 | Rectangular glass cell 10 x 5 Special optical glass with lid 320 -2500 nm |
| 1103-4015 | Rectangular glass cell 10 x 20 Special optical glass with lid 320 -2500 nm |
| 1103-4016 | Rectangular glass cell 10 x 50 Special optical glass with lid 320 -2500 nm |
| 1103-4017 | Rectangular glass cell 10 x 100 Special optical glass with lid 320 -2500 nm |
Rectangular Quartz Cells

Spectrosil Quartz, 190 to 2,700 nanometers. The only option here for measurements below 320 nanometers or above 2,500 nanometers.
| Part Number | Description |
|---|---|
| 1103-4002 | Rectangular quartz cell 10 x 2 Spectrosil Quartz with lid 190 -2700 nm |
| 1103-4003 | Rectangular quartz cell 10 x 5 Spectrosil Quartz with lid 190 -2700 nm |
| 1103-4004 | Rectangular quartz cell 10 x 10 Spectrosil Quartz with lid 190 -2700 nm |
| 1103-4005 | Rectangular quartz cell 10 x 20 Spectrosil Quartz with lid 190 -2700 nm |
| 1103-4006 | Rectangular quartz cell 10 x 50 Spectrosil Quartz with lid 190 -2700 nm |
| 1103-4007 | Rectangular quartz cell 10 x 100 Spectrosil Quartz with lid 190 -2700 nm |
Key Selection Criteria
Ten millimeters is the standard pathlength for routine UV-Vis and NIR absorbance work, and it’s the dimension most published methods and instrument software default to. JASCO’s rectangular cells span 1 to 100 millimeters, so shorter and longer pathlengths are available for concentrated or dilute samples. Match the pathlength in your method to the part number below before ordering.
Micro and Semi-Micro Cells

When sample volume is limited, the micro and semi-micro quartz cells below reduce internal volume without changing the pathlength options above. All six use the same Spectrosil Quartz as the standard quartz cells, so the wavelength range does not change: 190 to 2,700 nanometers.
Three of these cells (1103-0040, 1103-0043, and 1106-0041) use black-walled quartz. Black walls block stray light from reaching the detector, which matters most for low-absorbance samples, scattering samples, and fluorescence work, where light reflecting off clear cell walls would otherwise inflate the reading.
If you’re running fluorescence rather than absorbance, note that fluorescence detection sits at 90 degrees to the excitation beam, so it needs four polished cell faces. Every cell on this page has two.
| Part Number | Description |
|---|---|
| 1103-0036 | Micro rectangular quartz cell 2 x 5 Spectrosil Quartz with lid 190 -2700 nm |
| 1103-0037 | Micro rectangular quartz cell 2 x 10 Spectrosil Quartz with lid 190 -2700 nm |
| 1103-0039 | Semi-micro rectangular quartz cell 4 x 10 Spectrosil Quartz with lid 190 -2700 nm |
| 1103-0040 | Micro rectangular quartz black cell 2 x 10 Spectrosil Quartz with lid 190 -2700 nm |
| 1103-0043 | Semi-micro rectangular quartz black cell 4 x 10 Spectrosil Quartz with lid 190 -2700 nm |
| 1106-0041 | Micro short rectangular quartz black cell 2 x 10 Spectrosil Quartz with lid 190 -2700 nm |
Need Sub-Microliter Samples?
For nucleic acid and protein samples that don’t stretch to a full cuvette, JASCO’s One Drop – Low Volume accessory (SAH-769) measures as little as 0.6 microliters at a 0.2 millimeter pathlength, or 5 microliters at 1 millimeter in its standard configuration, on V-600 and V-700 Series spectrophotometers.
Verifying Photometric Accuracy
Every JASCO UV-Visible/NIR spectrophotometer ships with photometric accuracy validated against NIST SRM 930D, the National Institute of Standards and Technology’s reference material for spectrophotometer calibration. Published accuracy runs to ±0.0015 Abs between 0 and 0.5 Abs, and ±0.0025 Abs between 0.5 and 1 Abs. See the full specifications for wavelength accuracy and model-by-model detail.
If you run your own verification schedule, neutral-density glass filters are the established method for checking visible-region photometric accuracy, certified at 440, 465, 546.1, 590, and 635 nanometers.
Frequently Asked Questions
How do I clean and store a quartz or glass cuvette?
Rinse with a mild, non-abrasive detergent solution, then rinse several times with ultrapure water. Dry with clean, particle-free air or a volatile solvent such as ethanol. Avoid ultrasonic cleaners, which can pit polished optical surfaces, and never let cleaning solution sit and dry inside the cell. Keep the polished faces from touching metal, glass, or any other hard surface, and store cells in a protective case between uses
Can I run a short-pathlength cell in a standard 10 millimeter cuvette holder?
Yes, with a spacer sized to your cell’s pathlength. A properly fitted spacer centers the shorter cell in the holder’s optical path; without one, a 1 or 2 millimeter cell sitting loose in a 10 millimeter holder will give an inaccurate reading. Contact JASCO service and support to confirm spacer compatibility with your instrument model.