High Speed Separation of Disulfiram using UHPLC
January 5, 2024
Introduction
Disulfiram is a drug used to treat chronic alcoholism by producing an acute sensitivity to alcohol. We examined the utility of an X-Press PakC18S column (2.1mm I.D. x 50mmL.) packed with 2µm diameter packing material for the ultra-high speed separation of the above drug. The results were examined to determine whether the performance of the column and chromatography separation exceed those of conventional HPLC.
Experimental
The chromatography system utilized in this experiment was a JASCOX-LC system consisting of a 3185P UHPLC pump, 3080DG degasser, 3067CO column oven, 3070UV UV/Vis detector, Model 3059AS auto sampler and a chromatography data system.
Keywords
747007X
Results
Figure 1 shows the separation of a standard mixture of benzophenone (0.005mg/mL) and disulfiram (0.005mg/mL). The X-LC system provides an analysis time 7 times shorter than conventional HPLC while the resolution between the benzophenone and disulfiram was 5.4; the reproducibility of the peak ratio is 0.44%. These results well exceed those of conventional HPLC.
Featured Products:
High Speed Separation of Disulfiram using UHPLC
Introduction
Disulfiram is a drug used to treat chronic alcoholism by producing an acute sensitivity to alcohol. We examined the utility of an X-Press PakC18S column (2.1mm I.D. x 50mmL.) packed with 2µm diameter packing material for the ultra-high speed separation of the above drug. The results were examined to determine whether the performance of the column and chromatography separation exceed those of conventional HPLC.
Experimental
The chromatography system utilized in this experiment was a JASCOX-LC system consisting of a 3185P UHPLC pump, 3080DG degasser, 3067CO column oven, 3070UV UV/Vis detector, Model 3059AS auto sampler and a chromatography data system.
Keywords
747007X
Results
Figure 1 shows the separation of a standard mixture of benzophenone (0.005mg/mL) and disulfiram (0.005mg/mL). The X-LC system provides an analysis time 7 times shorter than conventional HPLC while the resolution between the benzophenone and disulfiram was 5.4; the reproducibility of the peak ratio is 0.44%. These results well exceed those of conventional HPLC.