Fluorescence Spectrometers Specifications

Spectrometer Specification

Model
FP-8200
FP-8300
FP-8500
FP-8600
FP-8700
Light Source
Continuous output Xe arc lamp with shielded lamp housing (150 W)
Light Source (for validation)
Integrated, selectable low pressure mercury lamp
Photometric System
Radio-photometer system using monochromatic light to monitor the intensity output of the Xe lamp
Monochromator
Holographic concave grating in modified Rowland mount
Wavelength Range (with Standard Detector)
Ex
Zero order, 200 - 750 nm
Zero order, 200 - 850 nm
Em
Zero order, 200 - 1010 nm
Zero order,
300 - 1400 nm
Wavelength Range (Optional)
Zero order, 200 - 900 nm
Zero order, 200 - 850 nm
N/A
option 1700nm
Automatic Cut Filters for High Order Diffraction
Option
Standard
Sensitivity *1 (RMS)
Base *2
4,500 : 1
8,000 : 1
8,500 : 1
3,500 : 1
1400 nm - 8,000:1
1700 nm - 1,000:1
Resolution
Ex
2.5 nm (at 546.1 nm)
1.0 nm (at 546.1 nm)
1.0 nm (at 546.1 nm)
1.0 nm (at 546.1 nm)
Em
2.5 nm (at 546.1 nm)
1.0 nm (at 546.1 nm)
1.0 nm (at 546.1 nm)
2.0 nm (at 546.1 nm)
Band Width
Ex
2.5, 5, 10, 20 nm
1, 2.5, 5, 10, 20 nm
1, 2.5, 5, 10, 20, L5, L10 nm
1, 2.5, 5, 10, 20, L5, L10 nm
Em
2.5, 10, 20, 40, L10, L20 nm
Wavelength Accuracy
Ex
±2.0 nm
±1.5 nm
±1.0 nm
±1.0 nm
Em
±2.0 nm
Wavelength Repeatability
Ex
±1.5 nm
±1 nm
±0.3 nm
±0.3 nm
Em
±0.6 nm
Wavelength Scan Speed
Ex
20, 50, 100, 200, 500, 1,000, 2,000, 5,000,
10,000, 20,000 nm/min
10, 20, 50, 100, 200, 500, 1,000, 2,000, 5,000, 10,000, 20,000,
60,000 nm/min
10, 20, 50, 100, 200, 500, 1,000, 2,000, 5,000, 10,000, 20,000, 60,000 nm/min
Em
20, 50, 100, 200, 500, 1,000, 2,000, 5,000, 10,000, 20,000, 60,000, 120,000 nm/min
Slew Speed
Ex
30,000 nm/min
60,000 nm/min
60,000 nm/min
Em
30,000 nm/min
120,000 nm/min
Response
20, 50, 100, 200,
500 msec, 1, 2, 4, 8 sec

10, 20, 50, 100, 200, 500 msec, 1, 2, 4, 8 sec
Detector
Ex: Silicon photodiode, Em: PMT
Ex: Silicon PD, Em: LN2 Cooled PMT
Photometric Range
-10,000 - 10,000
Sensitivity Selection
High, Medium, Low, Very Low, Manual, Auto SCS
Auto Gain
Standard
Shutter Function
Standard (Automatic control)
Sample Illuminating System
Horizontal illumination
Sample Compartment
10 mm rectangular cell holder, nitrogen purgeable
Recognition of IQ Accessory
Standard
Start Button
Standard
Analog Output
Standard
Instrument Communication
USB 2.0
Control and Data Processing
Spectra Manager™/CFR, iRM
Spectra Manager™/CFR
Spectral Correction
Option
Standard
(Spectral correction using a Rhodamine B ethylene glycol solution is standard; other jigs for spectral correction are available separately as options.)
Dimensions
490(W) × 545(D) × 270(H) mm
520(W) × 545(D) × 270(H) mm
570(W) × 545(D) × 270(H) mm
Weight
33.6 kg (74 lbs)
36 kg (79 lbs)
39 kg (86 lbs)
40 kg (88 lbs)
Power Requirement
270VA
Installation Environment
Temperature: 15 to 35°C, Humidity: Less than 85%

*1 : Minimum Signal-to-Noise ratio of Raman band of water, excitation 350 nm, band width Ex 5 nm Em 5 nm (FP-8600: Ex 5 nm Em 10 nm), response 2 seconds.
*2 : Noise is measured on the baseline.

Software Specification

A wide range of software applications and accessories can be used with Spectra Manager™ Suite.

Model
FP-8200
FP-8300
FP-8500
FP-8600
Data Station Type
Spectra Manager™/Spectra Manager™ CFR (Microsoft Windows® 7 or 8.1 Professional)
iRM Type
Handheld Intelligent Remote Module iRM-900
N/A
Measurement Programs
Spectra Manager™
Spectra measurement,

Quantitative measurement,

Fixed wavelength measurement,

Time course measurement,

3-D Spectra measurement,

Abs measurement
Spectra measurement,

Quantitative measurement,

Fixed wavelength measurement,

Time course measurement,

3-D Spectra measurement,

Abs measurement,

Phosphorescence measurement
Spectra Manager™ CFR *4
iRM-900 *5
Spectra measurement, Quantitative measurement, Fixed wavelength measurement, Time course measurement, 3D Spectra measurement, Abs measurement
N/A
Spectra Correction Program
Standard *6
Standard
Instrument Validation
Program and Hg lamp (Standard), Accessories (option)
Program and Hg lamp (Standard), Accessories (Standard)
Self Diagnosis
Standard
IQ Accessory Recognition and IQ Start
Standard

*4 : Spectra Manager™ CFR is the 21 CFR Part 11 compliant software package.
*5 : The iRM-900 handheld control module can be used with any PictBridge compatible printer.
*6 : Optional components are necessary.
Specifications are subject to change without notice.