Temperature Controlled Cell Holders - Peltier and Water

A range of temperature controlled cells holders recommended for kinetics and ramping to moderate temperatures

EHCS-760 Peltier Thermostatted Single Position Cell Holder (Air-Cooled)

This Peltier cell holder precisely maintains a specified sample temperature and can perform thermal ramping experiments up to 60 deg C.  A stirrer is included for uniform cell temperature. Air-cooled for temperature control (does not require a water circulator).

OPS-512 Optional Temperature Sensor
PS-512 OP sensor for direct temperature measurement in the sample cell.

Specifications

Thermostatted Cell10-mm rectangular cell (sample side)
Temperature control systemHeating/cooling system using Peltier effect
Heat radiating systemAir cooled
Stirring systemMagnetic stirrer (variable speed)
Temperature setting range5 to 70°C
Temperature control range10 to 60°C (for room temperature = 20°C)
Temperature control accuracy± 0.1°C
Temperature accuracy(Difference between displayed temperature and actual cell temperature)
20°C to 40°C ± 0.5°C
± 1°C when set temperature is outside the above range
Temperature setting difference(Difference between set temperature and displayed temperature)
20°C to 40°C ±0.3°C
± 0.5°C when set temperature is outside the above range


PSC-761/763 Automatic 6-position Peltier Cell Changer (Air-Cooled)

Multi-cell accessory for simultaneous temperature controlled measurement in up to 6 sample cells. Air-cooled Peltier effect to control sample temperature and ramping with high accuracy. Includes  stirrer.

OPS-513 Optional Temperature sensor
OPS-513 OP sensor for direct temperature measurement in the sample cell.

Specifications

Thermostatted CellUp to six 10 mm rectangular sample cells (ambient reference cell)
Temperature control systemHeating/cooling system utilizing Peltier effect
Heat radiating systemForced air cooling
Stirring systemMagnetic stirrer (variable speed)
Temperature setting range10 to 70°C
Temperature control range15 to 60°C (For room temperature = 20°C)
Temperature control accuracy±0.1°C
Temperature accuracy(Difference between the displayed temperature and temperature in the cell)
20°C to 40°C ±0.5°C
±1°C when the set temperature is outside the above range
Temperature setting difference(Difference between the set temperature and displayed temperature)
20°C to 40°C ±0.3°C
±0.5°C when the set temperature is outside the above range


STR-773 Water Thermostatted Cell Holder with Stirrer

Used with a temperature controlled circulating water bath (not supplied) to maintain a constant sample and reference cell temperature. The STR-773 accepts standard 10 mm path length cells. A magnetic stirrer is included to maintain temperature equilibrium in both cell positions.

 

Specifications

Cell10-mm cell
Operating temperatureRoom temperature plus 10°C to 90°C
Circulating water inlet tubeUni-tube (⌽6 x ⌽3 mm)


NCP-706 Water Thermostatted 6-position Automatic Cell Changer

Automatic 6-position cell changer for up to 6 cell positions (path lengths up to 10 mm). Single reference position. Temperature is controlled using a water-thermostatted circulator (not included).

 

 

Specifications

Cell MovementLinear
Cell Type10mm cell, Micro Cell (min. width 2mm)
Temperature control typeWater circulation
Operating temperatureRoom temperature plus 10° -90°C
Reference cell holderConstant temperature single-cell holder


CSP-909 Sample Holder Lid with Syringe Port Models

Optional  lid used for adding a reagent with a syringe without opening the sample chamber for starting fast reactions without a delay in measurement. The syringe port lid can be used with any single cell holders.

 

 

Specifications

Recommended syringe0507-0220 Micro syringe, 10 µL
0507-0246 Micro syringe, 100 µL

HMC-711 Water Thermostatted Micro Cell Holder (sample volume 100 µL, 200 µL)

Constant temperature single cell holder for micro cells. Accepts standard 10 mm path length cells with a minimum width of 2 or 4 mm. Used with a temperature controlled circulating water bath (not included).

 

 

Specifications

Compatible Cell5 or 10 mm path length rectangular cell
5 or 10 mm path length rectangular micro cell
Operating temperatureRoom temperature plus 10°C to 90°C
Circulating water tubeUni-tube (⌽6 x ⌽3 mm)

Optional cells

ItemP/NSpecification
Rectangular micro quartz cell1103-00255 mm path, 2 mm width
Rectangular micro quartz cell1103-003010 mm path, 4 mm width
Rectangular micro quartz cell1103-002610 mm path, 2 mm width
Rectangular micro quartz black masked cell1103-003110 mm path, 4 mm width
Rectangular micro quartz black masked cell1103-003210 mm path, 2 mm width


MHT-745 Manual 4-Position Water Thermostatted Turret Cell Holder

4-position turret cell holder can be rotated manually for measurement of up to four cells. Requires four 10×10 mm rectangular cells (not included). Requires a temperature controlled circulating water bath (not included)

 

Specifications

Thermostatted cells10-mm rectangular cell
Sample-side cell holder: 4
Reference side cell holder: 1
Operating temperatureRoom temperature plus 10 to 90°C
Circulating water inletUni-tube (⌽7 x ⌽5 mm)