EHCS-760 Peltier Thermostatted Single Position Cell Holder (Air-Cooled)
This Peltier cell holder precisely maintains a specified sample temperature and can perform thermal ramping experiments up to 60 deg C. A stirrer is included for uniform cell temperature. Air-cooled for temperature control (does not require a water circulator).
OPS-512 Optional Temperature Sensor
PS-512 OP sensor for direct temperature measurement in the sample cell.
Specifications
Thermostatted Cell | 10-mm rectangular cell (sample side) |
Temperature control system | Heating/cooling system using Peltier effect |
Heat radiating system | Air cooled |
Stirring system | Magnetic stirrer (variable speed) |
Temperature setting range | 5 to 70°C |
Temperature control range | 10 to 60°C (for room temperature = 20°C) |
Temperature control accuracy | ± 0.1°C |
Temperature accuracy | (Difference between displayed temperature and actual cell temperature) 20°C to 40°C ± 0.5°C ± 1°C when set temperature is outside the above range |
Temperature setting difference | (Difference between set temperature and displayed temperature) 20°C to 40°C ±0.3°C ± 0.5°C when set temperature is outside the above range |
PSC-761/763 Automatic 6-position Peltier Cell Changer (Air-Cooled)
Multi-cell accessory for simultaneous temperature controlled measurement in up to 6 sample cells. Air-cooled Peltier effect to control sample temperature and ramping with high accuracy. Includes stirrer.
OPS-513 Optional Temperature sensor
OPS-513 OP sensor for direct temperature measurement in the sample cell.
Specifications
Thermostatted Cell | Up to six 10 mm rectangular sample cells (ambient reference cell) |
Temperature control system | Heating/cooling system utilizing Peltier effect |
Heat radiating system | Forced air cooling |
Stirring system | Magnetic stirrer (variable speed) |
Temperature setting range | 10 to 70°C |
Temperature control range | 15 to 60°C (For room temperature = 20°C) |
Temperature control accuracy | ±0.1°C |
Temperature accuracy | (Difference between the displayed temperature and temperature in the cell) 20°C to 40°C ±0.5°C ±1°C when the set temperature is outside the above range |
Temperature setting difference | (Difference between the set temperature and displayed temperature) 20°C to 40°C ±0.3°C ±0.5°C when the set temperature is outside the above range |
STR-773 Water Thermostatted Cell Holder with Stirrer
Used with a temperature controlled circulating water bath (not supplied) to maintain a constant sample and reference cell temperature. The STR-773 accepts standard 10 mm path length cells. A magnetic stirrer is included to maintain temperature equilibrium in both cell positions.
Specifications
Cell | 10-mm cell |
Operating temperature | Room temperature plus 10°C to 90°C |
Circulating water inlet tube | Uni-tube (⌽6 x ⌽3 mm) |
NCP-706 Water Thermostatted 6-position Automatic Cell Changer
Automatic 6-position cell changer for up to 6 cell positions (path lengths up to 10 mm). Single reference position. Temperature is controlled using a water-thermostatted circulator (not included).
Specifications
Cell Movement | Linear |
Cell Type | 10mm cell, Micro Cell (min. width 2mm) |
Temperature control type | Water circulation |
Operating temperature | Room temperature plus 10° -90°C |
Reference cell holder | Constant temperature single-cell holder |
CSP-909 Sample Holder Lid with Syringe Port Models
Optional lid used for adding a reagent with a syringe without opening the sample chamber for starting fast reactions without a delay in measurement. The syringe port lid can be used with any single cell holders.
Specifications
Recommended syringe | 0507-0220 Micro syringe, 10 µL |
0507-0246 Micro syringe, 100 µL |
HMC-711 Water Thermostatted Micro Cell Holder (sample volume 100 µL, 200 µL)
Constant temperature single cell holder for micro cells. Accepts standard 10 mm path length cells with a minimum width of 2 or 4 mm. Used with a temperature controlled circulating water bath (not included).
Specifications
Compatible Cell | 5 or 10 mm path length rectangular cell 5 or 10 mm path length rectangular micro cell |
Operating temperature | Room temperature plus 10°C to 90°C |
Circulating water tube | Uni-tube (⌽6 x ⌽3 mm) |
Optional cells
Item | P/N | Specification |
---|---|---|
Rectangular micro quartz cell | 1103-0025 | 5 mm path, 2 mm width |
Rectangular micro quartz cell | 1103-0030 | 10 mm path, 4 mm width |
Rectangular micro quartz cell | 1103-0026 | 10 mm path, 2 mm width |
Rectangular micro quartz black masked cell | 1103-0031 | 10 mm path, 4 mm width |
Rectangular micro quartz black masked cell | 1103-0032 | 10 mm path, 2 mm width |
MHT-745 Manual 4-Position Water Thermostatted Turret Cell Holder
4-position turret cell holder can be rotated manually for measurement of up to four cells. Requires four 10×10 mm rectangular cells (not included). Requires a temperature controlled circulating water bath (not included)
Specifications
Thermostatted cells | 10-mm rectangular cell Sample-side cell holder: 4 Reference side cell holder: 1 |
Operating temperature | Room temperature plus 10 to 90°C |
Circulating water inlet | Uni-tube (⌽7 x ⌽5 mm) |